How to Calculate Leak Rate (SCCM)

A leak rate is expressed as a volume per unit of time. The rate is found by measuring the change in pressure multiplied by the volume. That number is then divided by the change in time multiplied by the surrounding atmospheric pressure to calculate the leak rate (SCCM).

Leak-rate-formula
ATM = Atmospheric pressure (PSIA)
V = Test volume (cm3)
Δp = The decay in pressure during test time
Δt = The amount of decay time (min.)
SCCM = Standard Cubic Centimeters per Minute

For example:
Leak rate = .02psi/0.05min * 50cm3/14.7psia
Leak rate = 0.4 * 3.401
Leak rate = 1.36 SCCM
 

Leak Rate Calculator

Leak Rate:

 
If you require assistance identifying any of these metrics or want to learn more about how to calculate SCCM for your leak detection application, call 801-264-1000 today. Our highly trained team will help you with calculations for determining SCCM.