Bombing Leak Test graph showing the test stages and reject limitZaxis Z-Charge and Z-Ion helium testing equipment

Helium Leak Test Method

Bombing Leak Test

Bombing leak testing exposes fine leaks in finished hermetic parts by driving helium into a sealed device, then measuring helium that escapes under vacuum.

Bombing, also called helium bomb testing, is a trace-gas method for finished hermetically sealed parts with no fill port. It can test sealed implantable devices, MEMS, relays and crystals after assembly is complete.

Parts soak in helium at elevated pressure inside a pressure vessel. If a leak path is present, helium enters the part; the part then moves to a vacuum chamber, where a helium mass spectrometer measures the gas as it bleeds back out.

That measurement reveals and quantifies leakage. Bombing reaches fine-leak sensitivity in the 10⁻⁹–10⁻¹⁰ mbar·l/s range.

Request for Quote

Leak Rate Calculator

How is Leak Rate Calculated →

What is a bombing leak test?

Bombing tests a finished sealed device without opening it or using a fill port. Helium is introduced through any existing leak path rather than a service connection.

The device receives a timed elevated-pressure helium soak in a pressure vessel. A leaking device retains helium, which can then be measured after transfer to a vacuum chamber.

The mass spectrometer measures helium released from the evacuated device and compares the signal with its reject limit. The result is a quantitative fine-leak decision in the 10⁻⁹–10⁻¹⁰ mbar·l/s range.

The station is configured for the part and its reject limit, then runs the soak, transfer and detector measurement as one recorded production sequence.

The pressure vessel begins the measurement. Elevated-pressure helium reaches any path between the sealed interior and the vessel, and the timed soak gives it a controlled opportunity to enter a leaking device. Because the gas travels through a leak path rather than a service connection, the part remains sealed throughout the test.

The vacuum measurement completes the cycle. In the evacuated chamber, helium retained by a leaking part can leave into the chamber and be measured by the mass spectrometer. Comparing that signal with the programmed reject limit gives a quantitative decision for the part.

Bombing covers fine leakage; a paired gross-leak check catches defects that release helium too quickly for the post-bomb measurement. Together they address both very small and very large leak paths in a finished hermetic device.

How a bombing test runs

1

BOMB

Place finished hermetic parts in a pressure vessel filled with helium at elevated pressure for a timed soak.

2

TRANSFER

Move parts from the bomb to the vacuum measurement chamber after the soak.

3

DETECT

A mass spectrometer measures helium bleeding back out of a leaking part.

4

RECORD

Compare the measured signal with the programmed reject limit and record the result.

A part within its programmed reject limit passes; a signal above the limit fails.

When bombing is the right method — and when it is not

Bombing is a fit when

  • ✓Finished hermetic parts with no fill port
  • ✓A need to verify fine-leak integrity after a part is sealed
  • ✓Fine/gross qualification that pairs a gross leak test with the bombing sequence

Sensitivity, cycle time and part volume

Bombing detects helium that entered a part during the pressure soak and escapes after the part reaches vacuum. The measurement is a total leak result rather than a leak-location result.

Gross leaks need their own screen: a large path can fill and empty fast enough for helium to escape before the vacuum measurement. Pairing bombing with a gross-leak test covers that behavior.

Soak time, transfer time and measurement time are controlled together. A repeatable cell coordinates the pressure vessel, vacuum measurement and gross-leak screen so every part follows the same cycle.

Pressure, soak duration and transfer time are controlled so helium exposure is comparable from part to part.

The mass spectrometer measures helium released in the evacuated chamber and compares the signal with the programmed reject limit.

Together, the fine-leak and gross-leak checks address the distinct behavior of small and large leak paths.

Bombing leak-test diagram showing helium pressure soak and vacuum measurement

Which Zaxis testers run bombing

Bombing is a helium method: the part is soaked in a pressure vessel, then measured under vacuum. Z-Charge handles the helium charge and Z-Ion the detection.

Applications for bombing testing

Implantable devices

Implantable devices

Finished implantable devices can be evaluated for fine-leak integrity after they are hermetically sealed.

MEMS and sealed sensors

MEMS and sealed sensors

Sealed MEMS, relays and crystals can be tested without a direct fill connection.

Hermetic electronics

Hermetic electronics

Finished hermetic electronic packages can be checked after sealing.

Bombing vs other leak test methods

Highlighted row is this method. ✓ yes · × no
MethodNon-destructiveNeeds an access portLocates the leakTypical use
Pressure decay✓✓×Sealed parts with a port
Vacuum decay✓✓×Parts and packaging
Mass flow✓✓×Flow-rate limits
Occlusion✓✓×Blocked passages
Burst×✓×Failure pressure
Pressure cracking✓✓×Valve opening pressure
Chamber✓××Sealed parts, no port
Sniffing✓✓✓Finding the leak site
Hard vacuum✓✓×Hermetic, safety-critical parts
Accumulation✓✓×Large parts, total leak rate
Bombing✓××Sealed parts, no fill port
Differential pressure✓✓×Large volumes, low leak rates
Trace gas✓✓✓Very small leaks, quantified

Bombing leak test FAQ

What is a bombing leak test?

It is a helium trace-gas method for finished hermetic parts with no fill port. Parts soak in helium under pressure, then a mass spectrometer measures helium released under vacuum.

What parts are suited to bombing?

Finished implantable devices, MEMS, relays and crystals are examples of parts suited to bombing.

Does bombing locate a leak?

No. It measures helium leaving the part to reveal and quantify total leakage; sniffing is used when the leak site must be located.

How sensitive is bombing?

Bombing provides fine-leak sensitivity in the 10⁻⁹–10⁻¹⁰ mbar·l/s range.

Why pair bombing with a gross leak test?

A large leak can fill and empty quickly, allowing helium to escape before the vacuum measurement. A gross-leak screen provides the paired coverage.

Does bombing require a fill port?

No. It is intended for finished hermetically sealed parts with no fill port.

What happens after the helium soak?

The part is transferred to a vacuum chamber, where a mass spectrometer measures helium bleeding back out.